Download user manual - PREVU Non-Invasive Cholesterol Test

Transcript
™
R
TRA
IN IN G
AN
See n
o
Secti
7
USER MANUAL
Q U A L I FI C AT
IO
ES
D
T O P E R AT O
PreVu Handheld Spectrophotometer
(English-Canada)
N
FO R
T
For use with the PreVu® Non-Invasive Skin Cholesterol Point of Care (POC) Test™
This User Manual is strictly intended for use by PreVu Test Operators in Canada
User Manual September 13 Final Version.indd 1
12-09-13 1:34 PM
PreVu Customer Service and Support
1-888-928-7205
www.prevu.com
PreVu Handheld Spectrophotometer
(PVU-SPECT-101-UNI)
P101-MA-025-01
This User Manual is strictly intended for use by PreVu Test Operators in Canada.
Manufactured by: Miraculins Inc., 6-1250 Waverley Street, Winnipeg, Manitoba, Canada, R3T 6C6
The PreVu® Non-Invasive Skin Cholesterol Test™ technology is protected by patents and patents pending in
North America and internationally.
PreVu® is a Registered Trademark of Miraculins Inc. All Rights Reserved. 2012.
™ Designates Property of Miraculins Inc. All Rights Reserved. 2012.
Wet-Nap® is a Registered Trademark of Nice Pak, Inc.
Copyright © 2012 by Miraculins Inc. All Rights Reserved. No part of this User Manual, including interior
design, cover design, and images, may be reproduced or transmitted in any form, by any means (electronic,
photocopying, recording, or otherwise) without the prior written permission of Miraculins Inc.
Printed in Canada.
User Manual September 13 Final Version.indd 2
12-09-13 1:34 PM
Table of Contents
1.
2.
3.
4.
5.
.
6.
Safety, Symbols and Classification............................................................................ 3
The PreVu® Non-Invasive Skin Cholesterol Point of Care (POC) Test........................... 5
The PreVu Handheld Spectrophotometer................................................................... 6
3.1 Spectrophotometer Specifications............................................................... 6
3.2 Spectrophotometer Orientation................................................................... 7
3.2.1 Exterior Features..................................................................... 7
3.2.2 The Latch Mechanism............................................................. 7
3.3 Preparing the Spectrophotometer................................................................ 8
3.3.1 Installing the Batteries ............................................................ 8
3.3.2 Charging the Batteries............................................................. 8
3.4 Turning the Spectrophotometer ON.............................................................. 8
3.5 Checking the Spectrophotometer after turning it ON.................................... 9
3.6 Navigation Buttons.................................................................................... 10
3.7 Turning the Spectrophotometer OFF.......................................................... 11
3.8 Setting the Language, Time, Date and LCD Contrast.................................. 11
3.8.1 Setting the Language............................................................ 12
3.8.2 Setting the Time and Date..................................................... 12
3.8.3 Setting the LCD Contrast....................................................... 13
3.8.4 Turning on the Qualification Mode......................................... 14
3.9 Spectrophotometer Intended Use Environments and Users........................ 15
Performing the PreVu Non-Invasive Skin Cholesterol Point of Care (POC) Test
using the PreVu Handheld Spectrophotometer........................................................ 16
4.1 Prepare Items Used for the Test................................................................. 16
4.2 Proceeding with the Test........................................................................... 17
Step 1: Calibration.............................................................................. 17
Step 2: Prepare Patient....................................................................... 18
Step 3: Apply Detector Reagent.......................................................... 19
Step 4: Apply PC Reagent................................................................... 19
Step 5: Blot Reagents ........................................................................ 20
Step 6: Apply Indicator Reagent.......................................................... 20
Step 7: Measurement......................................................................... 21
Step 8: Check Validity......................................................................... 21
Step 9: Clean Palm & Device.............................................................. 23
Reviewing Test Results........................................................................................... 24
5.1 Accessing Stored Results.......................................................................... 24
5.2 Interpretation of PreVu Test Results........................................................... 24
Guidelines for Optimum Test Performance............................................................... 25
6.1 Foam Pad Handling and Placement........................................................... 25
6.2 Test Operator and Patient Seating Positions – Dominant Hand Testing....... 26
6.3 Test Operator and Patient Seating Positions – Alternate Hand Testing........ 27
6.4 Reagent Dropper Bottles........................................................................... 28
6.5 Blotting..................................................................................................... 29
6.6 Reading the Skin Cholesterol Level........................................................... 30
www.prevu.com
User Manual September 13 Final Version.indd 1
1-888-928-7205 (Canada)
1
12-09-13 1:34 PM
7.
8.
9..
PreVu Test Operator Training and Qualification......................................................... 32
7.1 Training and Qualification Procedure......................................................... 32
7.1.1 Training Review........................................................................ 32
7.1.2 Practicing the Test.................................................................... 32
7.1.3 Qualification Testing................................................................. 32
7.1.4 Login....................................................................................... 32
7.1.5 Submit Results......................................................................... 33
7.1.6 Qualification Criteria................................................................. 33
7.1.7 Qualification Confirmation........................................................ 33
7.1.8 PreVu Test Operator Training and Qualification Summary.......... 33
.Solving Problems: Warnings and Error Messages..................................................... 34
Spectrophotometer Support and Care...................................................................... 37
9.1 Warranty and Repair Information............................................................... 37
9.2 Spectrophotometer Care........................................................................... 37
9.3 Cleaning Instructions................................................................................. 37
9.4 Alignment Shoe Replacement.................................................................... 37
2
User Manual September 13 Final Version.indd 2
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
SECT
SECTION 1
Safety, Symbols and Classification
International Compliance
This product complies with the requirements for Electromagnetic Compatibility for Medical
Electrical Equipment, IEC 60601-1-2, 2nd Edition, 2001 including Amendment 1/2004.
This product complies with the following electromagnetic emission standards:
ƒƒ RF Emissions: CISPR 11, Group 1, Class A
ƒƒ Voltage Fluctuations/Flicker Emissions: Per IEC 61000-3-3
This product complies with the following electromagnetic immunity standards:
ƒƒ Electrostatic Discharge (ESD), IEC 61000-4-2: Per IEC 60601 Test Level ±6 kV contact,
±8 kV air
ƒƒ Electrical Fast Transient/Burst, IEC 61000-4-4: Per IEC 60601 Test Level ±2 kV power
supply lines, ±1 kV input/output lines
ƒƒ Surge, IEC 61000-4-5: Per IEC 60601 Test Level ±1 kV differential mode, ±2 kV common
mode
ƒƒ Voltage Dips/Interruptions, IEC 61000-4-11: Per IEC 60601 Test Levels for typical
commercial or hospital environment
ƒƒ Power Frequency Magnetic Field, IEC 61000-4-8: Per IEC Test Level 3 A/m
When connected to a power adapter, the connected equipment must comply with FCC Part 15,
subparagraph B, Canadian Standard ICES-003, or international standard IEC 60950
or equivalents.
Symbols and Equipment Classifications
Manufacturer
This product has been tested by an authorized testing laboratory, and has been
found to meet the applicable standards in Canada and the United States
This product complies with the essential requirements of the relevant European
health, safety and environmental protection directives
Attention
Type BF Equipment
Device and batteries should be disposed of at a proper waste disposal site
www.prevu.com
User Manual September 13 Final Version.indd 3
1-888-928-7205 (Canada)
3
12-09-13 1:34 PM
Sterilization/Disinfection- Does not apply
Application Safety- Not suitable for use in the presence of flammable anaesthetic
CE Declaration
Manufacturer’s Name
Miraculins Inc.
Manufacturer’s Address
6-1250 Waverley Street
Winnipeg, MB, Canada, R3T 6C6
Model Name
PreVu Handheld Spectrophotometer
Model No.
101
Directive(s) Conformance
IVDD (98/79/EC)
European Authorized Representative
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands
ƒƒ Use Rechargeable Batteries Only (AA NiMH 2300 mAhr
minimum). Use of non-rechargeable batteries in this
instrument could cause the batteries to overheat and
damage the instrument.
ƒƒ Use UL-listed power adapter to charge the batteries in
this instrument (included with instrument).
ƒƒ Do not use this instrument to measure Skin
Cholesterol when connected to the power adapter.
ƒƒ No user serviceable parts in this product.
NOTE: This Spectrophotometer is intended to be used with the PreVu Non-Invasive Skin
Cholesterol Point of Care (POC) Test Reagent Kit only.
4
User Manual September 13 Final Version.indd 4
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
SECTION 2
The PreVu Non-Invasive Skin Cholesterol
Point of Care (POC) Test
For summary and explanation of the PreVu Non-Invasive Skin Cholesterol Point of Care (POC) Test
and principle of procedure, please refer to :
PRECAUTIONS
1. Dispense hanging drop of each reagent. In order
to prevent contamination, do not touch dropper
bottle nipple assembly to surface of skin.
• Section 4 of this User Manual
ENGLISH
™
PreVu Handheld Spectrophotometer
USER MANUAL
SUMMARY AND EXPLANATION
OF THE TEST
• The Package Insert included in the PreVu
Non-Invasive Skin Cholesterol Point of
Care (POC) Test Reagent Kit
(English-Canada)
The assessment of skin cholesterol allows for
the quantification of cholesterol present in the
tissues. Skin contains approximately 11%, by
weight, of all body cholesterol and has been cited
as “mirroring” vascular changes associated with
age and atherosclerosis1. A number of studies have
demonstrated a relationship between skin cholesterol
levels and CAD (reviewed in Lopukhin1) and more
recently studies showing a relationship between skin
cholesterol levels and treadmill stress test outcome2,
and skin cholesterol correlated with angiographically
proven CAD3 have been reported.
The PreVu POC Test affords a practical and effective
method of determining skin cholesterol. Skin
cholesterol as measured by the PreVu POC Test can
be used in conjunction with clinical evaluation and
other lipoprotein tests as part of risk assessment for
coronary artery disease.
PRINCIPLE OF THE PROCEDURE
The PreVu POC Test is a rapid two-step test for the
detection of skin cholesterol. The test is performed on
the palmar surface of the hand and requires no prior
fasting. The principle of the procedure is as follows:
For use with the PreVu® Non-Invasive Skin Cholesterol Point of Care (POC) Test™
This User Manual is strictly intended for use by PreVu Test Operators in Canada
• The Training Video CD included with the
PreVu Handheld Spectrophotometer
sa
ui
eV
Pr
mark of Miraculins Inc. 2
Trade
012.
ered
All
Rig
gist
Re
hts
Re
se
rve
d.
1. A solution (Detector) containing digitonin and
horseradish peroxidase enzyme (HRP) conjugated
to a copolymer is first placed on the palm of the
hand for one minute. During this time, digitonin
forms complexes with cholesterol in the skin. The
greater the amount of cholesterol present in the
epidermis, the greater the amount of Detector
will attach.
2. Unbound Detector is blotted off the palm with a
Blotting Stick.
3. A second solution, Indicator, is added to the test
sites for two minutes. The Indicator is a substrate
that reacts with HRP in the Detector to produce
a blue-green colour. The greater the amount of
Detector bound to skin cholesterol, the darker the
colour will become.
4. A hand-held spectrophotometer is used to
determine an epidermal skin cholesterol value
by measuring the hue† of the colour developed in
the test well.
Negative and positive controls applied to the palm in
parallel with the Detector solution serve as reagent
and procedural controls to monitor test performance.
MATERIALS
Includes:
• www.prevu.com
• Spectrophotometer
User Manual
Quick Reference Guide
• Training Video Presentation
English
Version 1.0
•
Materials Supplied in the PreVu POC Test Kit
(Contains 40 tests)
Alcohol Swab
Forty four (44) isopropyl alcohol (70% v/v) swabs
Foam Pad
Forty four (44) foam applicator pads with medicalgrade adhesive protected by a paper liner.
Wet-Nap® moist towelette
Training Video CD
Forty four (44) pre-moistened hand cleaning wipes.
Blotting Stick
for
The PreVu®
Non-Invasive Skin Cholesterol
Point of Care (POC) Test
Forty four (44) blotting sticks (cellulose fibre rods).
NOTE: Additional supply volumes provided for
practice/unintentional wastage.
Printed / Produced in Canada
One dropper bottle with 2.5 mL of digitonin
horseradish peroxidase conjugate in StabilZyme
SELECT® Stabilizer‡ as a buffer solution. Store
refrigerated. Stable at 2-8°C (36-46°F) for duration of
labelled expiry date.
P101-MA-002-00
2. Do not use kit components beyond expiry date
and do not mix reagents from kits with different
lot numbers.
3. Avoid exposure of reagents to excessive heat
or light.
INTENDED USE
The PreVu® Non-Invasive Skin Cholesterol Point
4. Observe good laboratory practices, i.e., do not
of Care (POC) Test (PreVu POC Test) is an in vitro
eat, drink, smoke or apply cosmetics in areas
diagnostic test for the quantification of skin
where biological reagents are used.
cholesterol. Like serum cholesterol, elevated skin
5. Bring reagents to room temperature (18-25°C or
cholesterol is associated with an increased risk of
64-77°F) prior to testing. Return all reagents to
coronary artery disease. Thus, skin cholesterol can be
refrigerator after use.
used as part of risk assessment for coronary artery
disease (CAD).
Detector (green cap)
PC - Positive Control (red cap)
One dropper bottle with 2.5 mL of digitonin
horseradish peroxidase conjugate in StabilZyme
SELECT® Stabilizer‡ as a buffer solution. Store
refrigerated. Stable at 2-8°C (36-46°F) for duration
of labelled expiry date. The PC solution is 10X more
concentrated than the Detector solution.
Indicator (blue cap)
One dropper bottle with 6.0 mL of 3, 3’, 5, 5’–
tetramethylbenzidine (TMB) – hydrogen peroxide
solution. Store refrigerated. Stable at 2-8°C (36-46°F)
for duration of labelled expiry date.
Materials Required But Not Provided
PreVu POC Handheld Spectrophotometer
(PVU-SPECT-101-UNI)
WARNINGS
Methylisothiazolone warning:
Methylisothiazolone in combination with
methylchloroisothiazolinone (MI/MCI) is routinely
used in the cosmetics industry as a biocide. MI/MCI
has been reported to be slightly irritating to the skin
at some concentrations. The PreVu POC Test reagents
contain only 100ppm of MI and are not expected
to cause skin irritation. However, use of the PreVu
POC Test should be discontinued in individuals who
experience skin irritation.
STORAGE INSTRUCTIONS
Store reagents between 2-8°C (36-46°F) for duration
of labelled expiry date.
4. In the clinical studies supporting the PreVu
POC Test safety and effectiveness only a small
number of African American subjects were
evaluated in relation to other patient subgroups.
Consequently, use of the PreVu POC Test for risk
assessment of CAD in African American subjects
must be carefully considered with all other risk
factors before any follow-up patient management
decisions are made.
5. Consult the PreVu POC Handheld
Spectrophotometer User Manual prior to use.
NOTE: It is recommended to test the hypothenar
eminence of the palm.
INTERPRETATION OF RESULTS
Skin cholesterol values less than 81 are below the
detection limit of the PreVu POC Test. Individuals with
skin cholesterol measurements equal to or below
134 are considered to be at normal risk of developing
CAD, whereas those with skin cholesterol values
greater than 134 are at increased risk (Table 1).
TEST PROCEDURE
Table 1. Skin Cholesterol Ranges
and Interpretation
Preparation for the Assay
1. Bring all reagents to room temperature (18-25°C
or 64-77°F).
2. Activate the PreVu POC Handheld
Spectrophotometer and consult User Manual.
3. Finish assay without interruption.
Assay Procedure
Follow test directions using the PreVu POC
Handheld Spectrophotometer.
1. Hands must be clean prior to the test. Wash
hands with soap and water or vigorously clean
outside portion of palm of one hand with WetNap® moist towelette provided. If using the moist
towelette, vigorously wipe the same area of the
palm with an Alcohol Swab, while the palm is still
wet from the Wet-Nap® wipe. Discard Wet-Nap®
moist towelette but save Alcohol Swab for use
after test. NOTE: Do not use other brands of
moist towelette to clean hands.
2. Air-dry palm.
7. Do not add any liquid to square well.
8. Wait 1 minute.
9. Slowly insert one end of the double-ended
Blotting Stick into the round, centre well of the
Foam Pad to absorb the Detector reagent. After
insertion press firmly and hold for 3 seconds
to ensure that all Detector reagent has been
absorbed, then invert Stick and insert the second
(clean) end into the same well (centre well) of
the Foam Pad. Press to make firm contact and
gently twist the Blotting Stick approximately one
quarter of a turn forward and then back while still
making contact with the palm. Once the centre
well has been blotted dry, remove the Blotting
Stick and slowly insert the second end into the
diamond-shaped well to absorb the PC reagent.
Discard the Blotting Stick. NOTE: Visually ensure
that Foam Pad and test wells are completely dry.
Test Validity
1. Liquid in square well is colourless.
2. Liquid in diamond well is blue-green.
Repeat test on other hand if assay is invalid. Consult
Technical Service at 1-888-928-7205 (Canada) if
assay fails the repeat test.
QUALITY CONTROL PROCEDURES
LIMITATIONS OF THE PROCEDURE
1. The wells in the Foam Pad must not be placed
over prominent creases on the palm as this
may allow the liquid to leak from the wells
during incubation.
2. Patients using topical medications and individuals
with skin disease (e.g. eczema, psoriasis) on the
hands should not be tested.
3. Patients who have recently used a dermatological
preparation, such as a hand lotion or barrier
cream, may produce aberrant skin
cholesterol values.
Normal range
of skin
cholesterol
Normal risk
Elevated skin
cholesterol
Increased risk
Normal risk
3. components of different lot numbers are mixed,
whether such components are purchased
individually or as part of a kit.
Patent Information
The PreVu Non-Invasive Skin Cholesterol test
technology is protected by patents and patents
pending in North America and Internationally.
Questions?
1-888-928-7205 (Canada)
www.prevu.com
Manufactured By:
EC Rep:
Miraculins Inc.
Emergo Europe
6-1250 Waverley Street Molenstraat 15
Winnipeg, Manitoba
2513 BH The Hague
Canada R3T 6C6
The Netherlands
Manufactured for Miraculins by Fisher Diagnostics,
part of Thermo Fisher Scientific.
PreVu® is a registered trademark of Miraculins Inc.
All rights reserved. 2011.
†
All colours can be characterized by hue (the
dominant shade), saturation (how much colour
of any hue is present) and lightness (the degree
of lightness or darkness of a particular colour)
in a three-dimensional space using cylindrical
coordinates. Lightness is the centre vertical axis
and saturation is the horizontal axis that extends
from the lightness axis. Hue is the angle at
which the saturation axis extends from the
lightness axis.
‡
StabilZyme SELECT® is an aqueous solution
containing a purified bovine protein and other
non-toxic stabilizing chemicals in MOPS
(3-(N-Morpholino) propanesulfonic acid), pH
6.2-6.7, buffer. The solution is preserved
with the following components: 0.02%
methylisothiazolone, 0.02% bromonitrodioxane
and 20 ppm Proclin 300 (Rohm and
Haas Company).
Table 2 shows the reproducibility of the PreVu POC
Test within-day, between days, and from batch
to batch.
Table 2. Reproducibility of PreVu POC:
Mean CV (range)
WithinDay
Day-toDay
Low skin
cholesterol
11%
(5-19%)
8%
(4-17%)
High skin
cholesterol
7%
(2-13%)
3%
(1-5%)
a
Batch-toBatcha
10%
(1-17%)
KIT CONTENTS
3 lots of PreVu POC were compared
(Contains 40 tests)
CLINICAL DATA
Skin cholesterol levels were measured using
the PreVu POC Test, in 750 individuals at three
angiography clinics (649 cases) and three out-patient
facilities (101 controls). Skin cholesterol levels were
significantly lower in control subjects compared to
the case population (105 vs. 123, p<0.001). The
distribution of controls and cases in the different skin
cholesterol ranges is shown below (Table 3).
Table 3. Distribution of Controls and
Angiography Clinic Patients by
Skin Cholesterol Range
Skin
Cholesterol
Range
Controls
(n=101)
Angiography
Patients
(n=649)
<81
12% (n=12)
10% (n=62)
81-134
79% (n=80)
56% (n=363)
>134
9% (n=9)
34% (n=224)
Table 4 shows the relationship between skin
cholesterol and CAD as determined by coronary
angiography. Skin cholesterol was correlated with
CAD even after adjusting for Framingham risk factors.
The strongest relationship was with multi-vessel
disease. This shows that skin cholesterol provides
new information about CAD risk that is independent
of Framingham risk factors: age, gender, serum LDL,
serum HDL, hypertension status, diabetes status, and
smoking status.
Instrument Calibration
When using the PreVu POC Handheld
Spectrophotometer, the instrument must be
calibrated prior to each test. The instrument will
prompt the operator to open the alignment shoe
completely and leave it open until a beep is heard. If
you have a tethered version of the spectrophotometer
please consult the instructions provided with that
device for the proper calibration procedure.
The instrument will then indicate whether calibration
was successful. If calibration was not successful,
the instrument will indicate to repeat the calibration
procedure. If the calibration error persists, please
refer to the User Manual.
CAD Risk
Category
Below limit of
detection
>134
2. the kit components are used beyond the
indicated expiry date, or
Precision
10. Add one drop of Indicator (blue cap) to all wells,
including the square well.
11. Wait two minutes.
12. Immediately read test well with the PreVu POC
Handheld Spectrophotometer.
Blot excess liquid and remove Foam Pad. Clean palm
with Alcohol Swab from Step 1.
Interpretation
<81
81-134
There are no conditions, warranties or
representations, expressed or implied, statutory
or otherwise, made by Miraculins Inc., its officers,
agents or distributors, regarding the performance of
this kit when:
1. the validated test procedure described in this
package insert is not followed,
SPECIFIC PERFORMANCE
CHARACTERISTICS
3. Remove adhesive backing from Foam Pad. Firmly
place Foam Pad on alcohol-wiped portion of hand
with the diamond well towards fingers.
4. Rest subject’s hand on flat surface.
5. Add one drop of Detector (green cap) to round
test well.
6. Add one drop of PC (red cap) to diamond well.
Skin
Cholesterol
Range
3. Sprecher DL, et al. Skin Tissue Cholesterol
(Skin Tc) is related to angiographically-defined
cardiovascular disease. Atherosclerosis 2003;
171:255-258.
Table 4. Relationship Between Skin Cholesterol
and CAD Determined by Coronary Angiography
(odds ratio, 95% confidence interval, p-value)
Parameter
Insignificant CADa
Any CADb
MultiVessel
CADc
Skin
1.06
1.07
1.08
cholesterol (0.99-1.12)
(1.01-1.13) (1.03-1.14)
(10 unit
0.06
0.02
0.004
increase)
Skin
cholesterol
1.05
1.06
1.08
Adjustedd (0.99-1.12) (1.01-1.13) (1.02-1.15)
(10 unit
0.13
0.03
0.009
increase)
a
b
c
d
≤ 50% stenosis in any coronary arteries
>50% stenosis in at least one coronary artery
>50% stenosis in more than one coronary artery
adjusted for Framingham Global Risk and trial site
REFERENCES
1. Lopukhin YuM. The skin and atherosclerosis
(a three-drop test). Physiochemical Aspects of
Medicine Reviews. 1992; 3:1-124.
2. Zawydiwski R, Sprecher DL, Evelegh MJ,
et al. Cholesterol 1,2,3™: A novel test for the
determination of skin cholesterol levels. Clinical
Chemistry 2001; 47:1302-1304.
1.
2.
3.
4.
5.
6.
7.
44 Alcohol Swabs – for external use only!
44 Foam Pads
44 Wet-Nap® moist towelettes
44 Blotting Sticks
Detector Solution (2.5 mL)
PC (This is the positive control solution) (2.5 mL)
Indicator Solution (6.0 mL)
NOTE: Additional supply volumes provided for
practice/unintentional wastage.
SYMBOL LEGEND
: Contains N Tests
: Store at
: Expiration Date
: Lot #
: Manufacturer
: In Vitro Diagnostic
: See instructions
: European Conformity
FRANÇAIS
UTILISATION VISÉE
Le test de cholestérol cutané non invasif PreVu®
Point of Care (POC) (test PreVu POC) est un test
diagnostique in vitro administré dans le but de
mesurer le cholestérol cutané. À l’instar du
cholestérol sanguin, les taux élevés de cholestérol
cutané sont associés au risque accru de
coronaropathies. En conséquence, il est possible
d’utiliser le test de cholestérol cutané pour évaluer le
risque de coronaropathies.
RÉSUMÉ ET EXPLICATION DU TEST
L’évaluation du cholestérol cutané permet de
quantifier le cholestérol présent dans les tissus.
La peau contient environ 11 % du cholestérol
de l’organisme, si on le mesure par le poids. On
semble croire que le cholestérol cutané reflète
les changements vasculaires liés à l’âge et à
l’athérosclérose1. Un bon nombre d’études montrent
un rapport entre les taux de cholestérol cutané et les
coronaropathies (étude de Lopukhin1). Des expertises
plus récentes soulignent un rapport entre les taux
de cholestérol cutané et les résultats des épreuves
d’effort sur tapis roulant2. En outre, une corrélation
entre le cholestérol cutané et les coronaropathies
confirmées par angiographie3 a été signalée.
Le test PreVu POC offre une méthode d’évaluation
pratique et efficace du taux de cholestérol cutané.
On peut utiliser les résultats du test PreVu POC en
association avec des évaluations cliniques et d’autres
épreuves de lipoprotéines pour mesurer le risque
de coronaropathies.
NOTE: For External Use Only!
The PreVu® POC Test
www.prevu.com
User Manual September 13 Final Version.indd 5
1-888-928-7205 (Canada)
5
12-09-13 1:34 PM
SECTION 3
The PreVu Handheld Spectrophotometer
3.1 Spectrophotometer Specifications
Electrical
Operating voltage (from AC Adapter)
5.0 VDC +/– 0.5 volts, 500 mA
Power Source
2 x AA Rechargeable Batteries
(NiMH 2300 mAhr minimum)
AC Power Adapter Accessory
UL-listed AC power adapter, 5 VDC 500 mA
UL Listed and CE-marked Power
Adapter Plugs
Phihong Model PSC03R-050 (North American,
European and United Kingdom Plugs included)
USB Cable
Mini-B to A-Type USB
Environmental/Misc.
Operating Temperature
10°C to 40°C (50°F to 104°F)
Operating Relative Humidity
30-70% (non-condensing)
Operating Atmospheric Pressure
700-1060 hPa
Shipping/Storage Temperature
-20°C to 80°C (-4°F to 176°F)
Shipping/Storage Relative Humidity
30-70% (non-condensing)
Shipping/Storage Atmospheric Pressure
500-1060 hPa
Mechanical
Dimensions
14.8 x 7.2 x 5.0 cm (5.8 x 2.8 x 2.0 in)
Weight
250 grams (0.5 pounds)
Shock and Vibration
Rugged, transportable
6
User Manual September 13 Final Version.indd 6
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
3.2 Spectrophotometer Orientation
3.2.1 Exterior Features
LED
Optical Aperture
Graphic LCD Screen
Navigation Buttons
Alignment Shoe
Alignment
Shoe Latch
Security Islet
Mini-USB Port
Battery Cover
3.2.2 The Latch Mechanism
In order to UNLATCH the Alignment Shoe, press the latch mechanism in from the left side of the
Spectrophotometer. The alignment shoe will then release. To latch the Alignment Shoe, close the
alignment shoe and while holding it in place, push the latch mechanism in from the right side of
the Spectrophotometer.
Closed Position
Open Position
Push Latch in to Release Alignment Shoe
www.prevu.com
User Manual September 13 Final Version.indd 7
1-888-928-7205 (Canada)
7
12-09-13 1:34 PM
3.3 Preparing the Spectrophotometer
3.3.1 Installing the batteries
Press in and push the battery cover towards the bottom of the Spectrophotometer. Place 2 type
AA NiMH rechargeable batteries (2300 mAhr minimum) in the battery compartment. Ensure the
batteries are installed as per the polarity indication in the battery compartment. Replace the
battery cover.
3.3.2 Charging the batteries
Batteries should be charged using the AC power adapter provided with the Spectrophotometer.
Connect the small (Mini-B) end of the USB Cable into the Mini-USB port on the bottom of the
Spectrophotometer. Connect the other end (A-type) to the UL Listed AC power adapter. There are
3 plugs included with the adapter: North American, European and United Kingdom. Select the
appropriate plug type and attach it to the adapter. Plug the adapter into the AC wall outlet.
Suitable batteries, UL listed adapter and USB cable are included.
Allow the Spectrophotometer to charge for at least 4 hours when installing new batteries. When
the instrument is turned off, and the batteries are charging, the LED will appear red. Please note
that the red LED will go out when the unit is turned on after charging is complete. If a low battery
warning is displayed on the Spectrophotometer, allow the Spectrophotometer to charge for at
least 15 minutes, or replace the batteries with 2 fully charged AA NiMH rechargeable batteries
(2300 mAhr minimum). Please refer to Section 8: Solving Problems: Warnings and Error
Messages for details.
3.4 Turning the Spectrophotometer ON
Press and hold OK (centre) button briefly to turn on
8
User Manual September 13 Final Version.indd 8
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
3.5 Checking the Spectrophotometer after turning it ON
The PreVu Handheld Spectrophotometer performs a Power-OnSelf-Test (POST) each time the instrument is turned on. If the
POST fails, the unit will display an error message. Please refer to
Section 8: Solving Problems: Warnings and Error Messages
for details.
If no errors are detected, the instrument will proceed to the
screen showing the PreVu logo.
The operating temperature of the Spectrophotometer must be
between 10°C and 40°C (50°F and 104°F). If the temperature of
the instrument is not within the specified range, it will display a
message indicating the current temperature of the instrument.
When the instrument warms or cools to a temperature in
the specified range, the startup procedure will automatically
proceed.
The Spectrophotometer will display a message reminding the
operator that the reagents must be at room temperature. For
details please refer to Section 4.1: Prepare Items Used for
the Test.
The STARTUP Screen will appear.
The battery charge status is shown in the top right corner of the
screen. When the battery is charging, the battery charge status
symbol will flash.
www.prevu.com
User Manual September 13 Final Version.indd 9
1-888-928-7205 (Canada)
9
12-09-13 1:34 PM
3.6 Navigation Buttons
Use the UP & DOWN
arrow buttons to scroll
through the menu
Press OK (centre)
to make selection
TIP: Holding OK for more than 3 seconds will turn the instrument off at any time.
Use the “NEXT” (RIGHT arrow) button to progress through each step of the PreVu Non-Invasive
Skin Cholesterol Point of Care (POC) Test procedure. Upon completion of tasks for the current
step, click the RIGHT arrow button to immediately go to the next step.
The “BACK” (LEFT arrow) button allows you to move backward a step.
10
User Manual September 13 Final Version.indd 10
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
This is useful if you accidentally selected forward or need to review previous instructions.
You should not need to use the back button because the test is designed to run in a forward
progression.
Some screens do not show a “BACK” button and thus you cannot go back to the previous step.
3.7 Turning the Spectrophotometer OFF
There are 3 options for turning the Spectrophotometer OFF:
1. The Spectrophotometer will automatically shut off after 5
minutes if the Spectrophotometer has not been used.
2. The Spectrophotometer can be manually turned off by
pressing and holding the OK button for 3 seconds at
anytime.
3. The Spectrophotometer can be turned off by selecting
SHUTDOWN on either the STARTUP or MENU screen.
3.8 Setting the Language, Time, Date and LCD Contrast
On the STARTUP screen, select MENU
On the MENU screen, select CONFIG (configuration)
www.prevu.com
User Manual September 13 Final Version.indd 11
1-888-928-7205 (Canada)
11
12-09-13 1:34 PM
3.8.1 Setting the Language
On the CONFIG screen, select LANGUAGE
Use the UP and DOWN arrows to highlight the appropriate
language. Press OK button to select the language.
3.8.2 Setting the Time and Date
NOTE: The Time and Date must be set each time the batteries are removed or replaced.
Setting the time
On the CONFIG screen, select TIME.
There are two clock settings: AM/PM and 24 hr
Use the RIGHT arrow button to highlight AM and then use the UP
and DOWN arrow buttons to select either AM (AM/PM) or 24 hr.
For the AM/PM clock, PM times can be selected by increasing
the hour past 12-noon.
Use the RIGHT or LEFT arrow button to select the hour. Use the
UP and DOWN arrow buttons to set the hour.
12
User Manual September 13 Final Version.indd 12
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
After setting the hour, press the RIGHT arrow button and use the
UP and DOWN arrow buttons to set the minute.
Follow the same procedure to set the second.
When the correct time is set, press the OK button to save
the changes.
Setting the Date
On the CONFIG screen, select DATE.
Use the UP and DOWN arrow buttons to select the day.
12
After setting the day, press the RIGHT arrow button and use the
UP and DOWN arrow buttons to select the month.
12
12
After setting the month, press the RIGHT arrow button and use
the UP and DOWN arrow buttons to select the year. When the
correct date is set, press OK button to save the changes.
3.8.3 Setting the LCD Contrast
On the CONFIG screen, select CONTRAST.
Use the UP and DOWN arrow buttons to darken and lighten the
contrast. When a desired contrast is achieved, press OK button to
save the setting.
www.prevu.com
User Manual September 13 Final Version.indd 13
1-888-928-7205 (Canada)
13
12-09-13 1:34 PM
3.8.4 Turning on the Qualification Mode
Turning on the Qualification Mode enables the PreVu Handheld Spectrophotometer to display
patient test scores on its screen as numerical values, to assist PreVu Test Operators during the
training and qualification stage. Please refer to Section 7: PreVu Test Operator Training and
Qualification for a full description of the training and qualification procedure.
Under Normal Operating (default) Mode settings, the PreVu Handheld Spectrophotometer displays
patient test results by category, relaying a Normal, Elevated or Borderline Skin Cholesterol
level. This is based upon a patient’s test score falling within a defined numerical range, which is
measured and recorded by the instrument and then translated into a category description for the
screen display. Patients are then accordingly advised what category they are in and the results
are interpreted by the PreVu Test Operator (see Section 5.2: Interpretation of Test Results).
Patients are not provided with their individual numerical test scores as a matter of course – only
the range they have tested within - as it is the category placement overall that determines
potential next-steps on a clinical level.
However, during the training and qualification stage, PreVu Test Operators do require access
to a patient’s individual numerical test score in order to enter this data at www.prevu.com to
demonstrate that they are able to obtain correct Skin Cholesterol results, as determined by
comparison to established references values. The series of numerical test scores entered online
by the PreVu Test Operator are then automatically calculated to determine PreVu Test Operator
proficiency and their qualification to begin testing actual patients.
The steps below outline the procedure for temporarily turning on the Qualification Mode of the
PreVu Handheld Spectrophotometer so that individual, numerical test scores (i.e. 95, 82, 115)
display on the Spectrophotometer screen in place of the category test results (Normal, Elevated,
Borderline). PreVu Test Operators can then enter this data at www.prevu.com under the PreVu
test Operator Training and Qualification section of the web site, when doing qualification testing
on their final three test volunteers.
When the PreVu Handheld Spectrophotometer is turned off or shuts down on its own, it defaults
back to Normal Operating Mode displaying the category test results.
On the STARTUP screen, select MENU
On the MENU screen, select Qualification
14
User Manual September 13 Final Version.indd 14
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
On the QUALIFICATION screen use the Up or Down arrow button
to highlight the Qualification Mode setting as being ON or OFF.
Press OK button to save the setting.
SKIN CHOLESTEROL
In Qualification Mode, the RESULT screen will display the
patient’s individual, numerical test score.
3.9 Spectrophotometer Intended Use Environments and Users
The PreVu Handheld Spectrophotometer is intended for use with the PreVu Non-Invasive
Skin Cholesterol Point of Care (POC) Test in environments usually including, but not limited to
doctors’ offices, nursing stations, pharmacies, health and fitness clubs, employee and corporate
health assessment centres, mobile health assessment units, community health fairs and other
permanent and temporary clinical service settings accessible by the general public.
The PreVu Handheld Spectrophotometer is intended to be used by any individual who has been
trained and qualified in its proper operation for use with the PreVu Non-Invasive Skin Cholesterol
Point of Care (POC) Test (see Section 7: PreVu Test Operator Training and Qualification), for
testing in environments as referred to above. Such individuals would usually include, but not
be limited to doctors, nurses, doctors’ assistants, lab technicians, pharmacists, pharmacist
technicians, dieticians, general staff and other individuals both with and without formal
medical training.
www.prevu.com
User Manual September 13 Final Version.indd 15
1-888-928-7205 (Canada)
15
12-09-13 1:34 PM
SECTION 4
Performing the PreVu Non-Invasive Skin
Cholesterol Point of Care (POC) Test Using
the PreVu Handheld Spectrophotometer
NOTE: Read this section and the PreVu Non-Invasive Skin Cholesterol Point of Care (POC)
Test Reagent Kit Package Insert carefully before testing. Also view the Training Video CD or
the Training Video presentations at www.prevu.com.
4.1 Prepare Items Used for the Test
The following items should be readily available:
ƒƒ At least 1 absorbent tissue will be needed for each test (not supplied in Reagent Kit).
Ensure that they are non-lotion, lint-free tissues.
ƒƒ Wet-Nap® moist towelette
ƒƒ Alcohol swab
ƒƒ Foam test pad
ƒƒ Reagents
ƒƒ Blotting stick
Bring the reagent bottles to room temperature (approximately 18 to 25°Celsius or 64 to
77°Fahrenheit)
Remove reagents from kit and bring to room temperature at least 30 minutes prior to
testing patient. Once the testing day is completed the reagents need to be returned to the
refrigerator for overnight storage.
When the PreVu Handheld Spectrophotometer is turned ON, a
reminder screen will appear for 5 seconds.
Set up for test as follows:
1. Remove caps on all bottles and line the bottles up in the following order:
Detector, PC (Positive Control) and Indicator... or Green, Red and Blue by bottle cap
and label colours shown. This is the order they will be used in the test.
2. Place one absorbent tissue in front of the patient.
3. Have available a Wet-Nap® moist towelette, alcohol swab, foam pad and
blotting stick.
16
User Manual September 13 Final Version.indd 16
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
4.2 Proceeding with the Test
When the STARTUP screen appears, select BEGIN TEST.
NOTE: Also review Section 6: Guidelines for Optimum Test Performance in addition to the
steps that follow below.
Step 1: CALIBRATION
Prior to each test, the instrument must be calibrated.
CALIBRATE: If the shoe is locked shut, unlatch the shoe, and then
open the shoe completely.
When the shoe is fully open, the CALIBRATING screen will
appear. It is important that the shoe remains fully open until the
calibration is complete.
When calibration is complete, you will hear a short beep, and if
calibration was successful, the screen will show SUCCESSFUL.
If calibration was not successful , the instrument will indicate to
repeat the procedure.
Please refer to Section 8: Solving Problems: Warnings and Error Messages if the calibration
error persists.
TIP: Place Spectrophotometer back on table after calibration.
www.prevu.com
User Manual September 13 Final Version.indd 17
1-888-928-7205 (Canada)
17
12-09-13 1:34 PM
Step 2: PREPARE PATIENT
Perform the following steps to properly clean the patient’s hand and apply the foam pad.
CLEAN PALM
ƒƒMake sure that the patient washes their hands vigorously with
soap and water OR vigorously clean the outside portion of palm
of one hand with the Wet-Nap® moist towelette that
is provided. NOTE: Do not use other brands of moist towelette
to clean palm.
ƒƒ Dry the palm using a clean paper towel or other absorbent,
lint-free tissue.
ƒƒ Wipe the outside portion of the palm of the hand vigorously
with an alcohol swab and wait about 30 seconds for the
cleaned area to air dry. Keep the alcohol swab for use after
the test.
NOTE: Palm must be completely dry before proceeding with the next step.
PLACE FOAM PAD & SECURE
ƒƒ Remove protective liner from the foam pad and firmly place the
pad on the hypothenar eminence where the palm was cleaned
with the alcohol swab (see diagram above).
ƒƒ You will note the foam pad contains three different shaped
wells: square, round, and diamond.
ƒƒ Place the diamond well toward the fingers as shown in the
screen image above.
ƒƒ Avoid placing the foam pad on prominent creases of the palm.
Ask patient to invert palm on absorbent tissue and firmly press
down to secure pad. Ensure the foam pad is firmly stuck to the
palm by checking the edges of the foam pad.
ƒƒ Press RIGHT arrow button to proceed to next screen.
TIP: Use the palm of the hand of the patient that is the same as your dominant hand for
ease in performing the test (i.e. use the patient’s right hand if you are right handed).
NOTE: Full test procedure can be viewed on the Training Video CD and at www.prevu.com.
The PreVu POC Test Reagent Kit contains enough reagent for 40 tests, plus some additional
reagent for unintentional wastage. Keeping count of the number of tests performed will
help PreVu Test Operators know when they are coming to the end of their reagent supply.
18
User Manual September 13 Final Version.indd 18
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
Step 3: APPLY DETECTOR REAGENT
Green LED
One drop of the Detector (green cap) must be applied to the
round test well (centre well). The LED to the top left of the
screen will appear green (corresponding to the green cap of
the Detector).
It is important that only one drop be added to the proper
well. Apply the Detector by hovering the tip of the dropper
bottle over the well without touching it and then dispensing
one drop cleanly into the well.
NOTE: Press RIGHT arrow button IMMEDIATELY after addition of Detector to proceed to next
screen. It is important that the Detector has exactly one minute to react before it is removed.
TIP:
ƒƒ Angle reagent bottle at 45°.
ƒƒ Do not touch the skin with bottle tip.
ƒƒ If any liquid ends up on the foam pad, dry it off gently with a clean
absorbent tissue.
ƒƒ Check to confirm there has been no reagent leakage around the base
of the foam pad where it meets the skin. If there is leakage at any point
during the test, discontinue the test immediately.
ƒƒ Do not test the same hand again for a 24-hour period. You must use the
patient’s opposite hand for the second test within 24 hours
ƒƒ .
Red LED
Step 4: APPLY PC REAGENT
Immediately after screen appears add one drop of Positive
Control (PC) (red cap) to the diamond shaped well. The LED
to the top left of the screen will appear red (corresponding to
the red cap of the PC Reagent).
NOTE: It is important that NO liquid should be added to the square well at this time.
The hand should be held still for the ensuing one minute incubation period. The PC Reagent
screen will appear for a maximum of 10 seconds. If you have completed the PC step earlier
than the 10 seconds, the screen can be advanced by pressing the RIGHT arrow button.
White/Blue LED
The LED light on the top left of the screen will flash white/
blue during the wait time.
After 10 seconds, or if the screen is advanced manually,
the timing bar will display on the screen. The number to the
bottom right of the bar indicates the number of seconds
remaining. The LED will flash white/blue until the reaction
time has expired.
“PREPARE”
When the Detector has been incubating for 50 seconds (10
seconds remaining), you will hear a short beep, and you will
be prompted to PREPARE TO BLOT.
www.prevu.com
User Manual September 13 Final Version.indd 19
1-888-928-7205 (Canada)
19
12-09-13 1:34 PM
Step 5: BLOT REAGENTS
After the one minute timer expires, you will hear a long beep. You
will be prompted to “BLOT REAGENTS”.
Firmly blot the liquid from the Detector (round) and PC (diamond)
wells of the foam pad.
Slowly insert one end of the double-ended blotting stick into the round centre well of the foam
pad to absorb the Detector Reagent.
After insertion press firmly and hold for 3 seconds to ensure all excess Detector Reagent has
been absorbed then invert blotting stick and insert second (clean and unused) end into the same
round well (centre well) of the Foam Pad.
Press to make firm contact and gently twist the blotting stick approximately one quarter of a turn
forward and then back while still making contact with the palm. Once the round centre well has
been blotted dry, remove the blotting stick and slowly insert the second end into the diamondshaped well to absorb the excess PC Reagent. Discard the blotting stick.
Press RIGHT arrow button to proceed to next screen.
CAUTION: Blot round (centre) well first, blot diamond well second.
Visually inspect wells to ensure all liquid has been removed. It is important to remove ALL liquid
from BOTH the centre (round) and diamond wells.
NOTE: Blotting directions also presented on the Training Video CD and at www.prevu.com
and in Section 6: Guidelines for Optimum Test Performance.
Step 6: APPLY INDICATOR REAGENT
Apply one drop of Indicator (blue cap) to ALL three wells in the following order: square,
round, diamond.
Blue LED
The LED to the top left of the screen will appear blue
(corresponding to the blue cap of the Indicator Reagent).
NOTE: After applying the Indicator to all wells, you must
IMMEDIATELY proceed to the next step by pressing the
RIGHT arrow button.
White/Blue LED
“PLEASE WAIT”
After the Indicator has been applied to all three wells, the
patient must hold their hand steady and level for 2 minutes (120
seconds). The timing bar, and the number of seconds remaining
will appear on the screen.
The LED to the top left of the screen will flash white/blue during
this time. It is important that the solutions incubate for exactly
two minutes before the measurement is taken.
20
User Manual September 13 Final Version.indd 20
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
White/Blue LED
“PREPARE TO READ”
When 10 seconds remain, you will hear a short beep, and the
screen will ask you to OPEN SHOE & ALIGN ON PAD.
p. You
mond)
“ALIGN ON PAD”
Hold the shoe of the Spectrophotometer perpendicular to
the foam pad. Position the Spectrophotometer such that the
rectangular inset on the Spectrophotometer shoe fits securely
around the foam pad with the target window on the plastic shoe
directly over the round well in the centre of the foam pad.
am
as
same
a turn
has
nd-
Step 7: MEASUREMENT
Just prior to the end of 2 minutes, you will hear two short and
one long beep, and the screen will show PRESS DOWN & HOLD.
Depress and hold the instrument over the centre well until the
instrument shows MEASUREMENT COMPLETE.
quid
m
TIP: Do not lift the device off the palm until the screen changes to
“MEASUREMENT COMPLETE”.
It is IMPORTANT that the measurement is taken IMMEDIATELY
after the 2 minute timer expires. If the measurement is not taken
until 2 minutes and 10 seconds (the very outside incubation
range), the screen will show INVALID TEST and it will ask you
to REPEAT the TEST, which then must be done on the patient’s
other palm if conducted within a 24-hour period.
Step 8: CHECK VALIDITY
You must visually confirm if the test is valid or not by looking at
the outer control wells (square and diamond).
(120
ining
uring
ctly
www.prevu.com
User Manual September 13 Final Version.indd 21
1-888-928-7205 (Canada)
21
12-09-13 1:34 PM
The test is considered VALID when it meets the following criteria:
1. The liquid in the square well is colorless
2. The liquid in the diamond well is blue/green
VALID TEST RESULT
If the test does meet the two criteria defined above, press the
(YES) RIGHT arrow button to indicate that the test is VALID.
Upon completion of a valid test, the Skin Cholesterol Level will
appear on the RESULT screen.
SKIN CHOLESTEROL
NORMAL
12
On the RESULT screen, you will see:
ƒƒ Skin Cholesterol Level
ƒƒ Unique 4 digit ID
ƒƒ Date the test was performed
When you are finished with the current test result, and have recorded down the data showing on
the screen, press the (NEXT) RIGHT arrow button to continue.
INVALID TEST RESULT
If the test is NOT VALID according to the criteria defined above,
press the (NO) LEFT arrow button.
To be completely sure, the screen will prompt you to CONFIRM
VALIDITY; IS THE TEST VALID? Reinspect the wells of the foam
pad and answer the question on the screen. If the test does not
meet the 2 criteria defined above, press the (NO) LEFT arrow
button to confirm that the test is NOT VALID.
(Note: If the test is VALID, please press the (YES) RIGHT arrow
button to proceed to VALID TEST RESULT Section above.)
The INVALID TEST screen will appear, asking you to REPEAT
TEST. When OK is pressed, the screen will prompt you to
CALIBRATE the instrument and then prepare the patient for the
next test.
22
User Manual September 13 Final Version.indd 22
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
Step 9: CLEAN PALM & DEVICE
You MUST clean the patient’s hand and the instrument
(Alignment Shoe) whenever you complete a valid test or abort a
test after reagents have been applied to the patient. Use the
alcohol swab you saved at the beginning of the test and an
absorbent tissue to clean and dry the palm, and an absorbent
tissue to dry the Alignment Shoe.
NOTE: If you will be immediately performing another test on the same patient, you must
use their other hand OR wait 24 hrs before performing another test on the same hand.
Cleaning up the Patient
To ensure the patient is properly cleaned after a test is completed (or aborted), perform the
following steps:
ƒƒ Invert palm on absorbent tissue to remove liquid from wells.
ƒƒ Peel off pad.
ƒƒ Clean palm with alcohol swab saved from the beginning of the test.
ƒƒ Dry with absorbent tissue.
Cleaning up the Spectrophotometer
To avoid contaminating future tests, it is important that you properly clean your instrument after
a test is completed (or aborted). To remove any chemical residue from the previous test, simply
clean your instrument (Alignment Shoe) with a dry absorbent tissue.
Please refer to Section 9: Spectrophotometer Support and Care.
WARNING: DO NOT allow any liquid into the Spectrophotometer during cleanup, especially
the Optical Aperture (see page 10).
The screen will remain on the instrument for 5 seconds, or
alternatively, it can be advanced by pressing the (NEXT) RIGHT
arrow button.
You are now ready to perform the next PreVu Non-Invasive Skin
Cholesterol Point of Care (POC) Test.
www.prevu.com
User Manual September 13 Final Version.indd 23
1-888-928-7205 (Canada)
23
12-09-13 1:34 PM
SECTION 5
Reviewing Test Results
5.1 Accessing Stored Results
It is possible to recall the 10 most recent test results. To view the
results, select REVIEW from the MENU screen.
SKIN CHOLESTEROL
ELEVATED
The screen will show the result of the most recent test
performed. The screen will display the unique 4-digit ID #, the
date and time the test was performed, and the Skin
Cholesterol Level.
The previous test results (category format only: Normal, Elevated
and Borderline), can be reviewed by pressing the UP arrow
button.
To return to the MENU screen, press the OK button.
In Qualification Mode, while numerical test scores are displayed on screen during the test for the
benefit of PreVu Test Operators in training, numerical test scores cannot be accessed afterwards
under REVIEW. PreVu Test Operators are advised to record numerical test scores at the time of
the test to enter them online for qualification at www.prevu.com, please refer to Section 7: PreVu
Test Operator Training and Qualification.
5.2 Interpretation of PreVu Test Results
The degree of colour hue measured during the PreVu POC Test is proportional to the amount of
bound Skin Cholesterol on the surface of the skin, and places the patient on a scale showing
their risk of coronary artery disease based on the clinical data for the biomarker.
A Skin Cholesterol value of 110 is chosen (based
on clinical studies) as the value at which patients
begin to acquire an increased burden of risk of
coronary artery disease.
The PreVu Handheld Spectrophotometer displays
patient test results as being either Normal,
Elevated or Borderline based upon a patient’s
reading falling within a defined numerical range,
which the instrument measures and records
internally.
Elevated Skin Cholesterol Level
PreVu® Score 110 or higher
110
Elevation correlates with an increased
risk of Coronary Artery Disease
Borderline Skin Cholesterol Level
PreVu® Score between 100 – 109
100
The adjacent table presents an outline of the three
categories of PreVu Test results.
Normal Skin Cholesterol Level
PreVu® Score below 100
Skin Cholesterol risk marker is within normal range
For more information on how these categories are to be interpreted for patients, please visit
www.prevu.com.
24
User Manual September 13 Final Version.indd 24
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
SECTION 6
Guidelines for Optimum Test Performance
This section presents a series of guidelines for best test performance, which will be of assistance
in helping a PreVu Test Operator become more comfortable with conducting the PreVu NonInvasive Skin Cholesterol Point of Care (POC) Test. It will also assist in helping you complete the
training and obtain qualification (see Section 7: PreVu Test Operator Training and Qualification),
which is required prior to being able to test an actual patient.
6.1 Foam Pad Handling and Placement
ƒƒ When peeling the protective backing off the foam pad and placing the pad onto the
patient’s hand, the PreVu Test Operator must take care not to come into contact with the
adhesive backing.
ƒƒ Peel back a portion of the foam pad, positioning the square well towards the patient,
and place this portion on the patient’s hand first, removing the balance of the protective
backing while at the same time laying down the remainder of the foam pad.
ƒƒ Ensure the foam pad is laid down on the palm without being stretched or pulled from end
to end.
ƒƒ Do not place the wells of the foam pad over creases on the palm of the hand. Creases on
the palm of the hand increase the chance of reagent leakage along the edges of the pad
which will result in an Invalid Test result.
ƒƒ Place the foam pad on the hypothenar eminence of the palm as indicated in the
below diagram, with the diamond well towards the fingers as shown in the image below.
Hypothenar Eminence
ƒƒ The test should be performed with appropriate overhead lighting to ensure that the visual
reading of the negative (square) and positive control (diamond) wells of the foam pad is
carried out accurately.
www.prevu.com
User Manual September 13 Final Version.indd 25
1-888-928-7205 (Canada)
25
12-09-13 1:34 PM
6.2 Test Operator and Patient Seating Positions – Dominant Hand Testing
ƒƒ The test should be performed seated at a table and the PreVu Test Operator should be
positioned at the right or left corner of the table with the subject seated on the
adjacent side.
ƒƒ The PreVu Test Operator should hold the PreVu Handheld Spectrophotometer in his/her
dominant hand and test the same hand on the patient. In this manner, if you are right
handed you should hold the Spectrophotometer in your right hand and test the patient’s
right hand. If the left hand is your dominant hand, you should hold the Spectrophotometer
in your left hand and test the patient’s left hand. This is to ensure the best positioning of
the Spectrophotometer possible to take the skin cholesterol reading, and a steady hand to
achieve a stable and level fit of the Alignment Shoe over top of the foam pad.
ƒƒ The below diagrams show the optimum PreVu Test Operator and patient seating positions
for a right-handed operator (dominant hand), seated at both a right corner table and left
corner table configuration and testing the patient’s right hand. In all diagrams, please note
the location of the foam pad on the patient’s hypothenar eminence (represented by the
black strip), and the Spectrophotometer held in the PreVu Test Operator’s dominant hand.
RIGHT HANDED OPERATOR
RIGHT HANDED OPERATOR
Dominant hand testing
Dominant hand testing
Patient
Patient
PreVu Test Operator
PreVu Test Operator
ƒƒ The below diagrams show the optimum PreVu Test Operator and patient seating positions
for a left-handed operator (dominant hand), seated at both a right corner table and left
corner table configuration and testing the patient’s left hand.
LEFT HANDED OPERATOR
LEFT HANDED OPERATOR
Dominant hand testing
Dominant hand testing
Patient
Patient
PreVu Test Operator
26
User Manual September 13 Final Version.indd 26
PreVu Test Operator
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
ƒƒ Avoid seating the patient across the table as demonstrated in the diagram below. This
could necessitate stretching and limit dexterity and balance during the blotting and
reading portions of the test and result in an Invalid Test result.
Patient
PreVu Test Operator
6.3 Test Operator and Patient Seating Positions – Alternate Hand Testing
ƒƒ If the first test conducted is determined to be an Invalid Test result and you wish to run
a second test immediately, you must conduct that second test on the patient’s opposite
hand. The same hand may not be tested again for a 24 hour period due to the risk of
reagent residue on the palm from the first test interfering with the second test reading.
ƒƒ To ensure that the correct test procedure continues to be followed (the PreVu Test
Operator continuing to use his/her dominant hand in holding the PreVu Handheld
Patient
Spectrophotometer and the skin cholesterol reading not being taken across the palm from
LEFT HANDED OPERATOR
the thumb side, (see Section 6.6 Reading the Skin Cholesterol Level)) the PreVu Test
Operator and the patient must change positions.
Alternate hand testing
ƒƒ The diagram below shows the best repositioning for a right-handed PreVu Test Operator
and patient, seated at both a right corner table and left corner table configuration, to
conduct an alternate hand test (testing the patient’s left hand). This position is not as
natural as the position for a dominant hand test, but it will still allow for the correct
PreVu Test Operator
procedure to be followed.
PreVu Test Operator
Patient
RIGHT HANDED OPERATOR
RIGHT HANDED OPERATOR
Alternate hand testing
Alternate hand testing
Patient
PreVu Test Operator
www.prevu.com
User Manual September 13 Final Version.indd 27
1-888-928-7205 (Canada)
27
12-09-13 1:34 PM
ƒƒ The diagram below shows the best repositioning for a left-handed PreVu Test Operator
and patient, seated at both a right corner table and left corner table configuration, to
conduct an alternate hand test (testing the patient’s right hand). This position is not
as natural as the position for a dominant hand test, but it will still allow for the correct
procedure to be followed.
Patient
PreVu Test Operator
LEFT HANDED OPERATOR
LEFT HANDED OPERATOR
Alternate hand testing
Alternate hand testing
PreVu Test Operator
Patient
6.4 Reagent Dropper Bottles
ƒƒ The reagents are provided in dropper bottles with a special tip to control the flow of
reagent to the precise amount needed for the test, which is achieved by firmly yet slowly
squeezing the bottle between the thumb and index finger until the drop builds and cleanly
detaches from the dropper bottle tip.
ƒƒ The PreVu Test Operator should practice squeezing out several drops of reagent onto a
paper towel to get a feel for the squeezing pressure required to generate full, clean drops.
ƒƒ If the dropper bottle tip appears wet or has a small droplet at the tip prior to the test,
dry the tip with an absorbent, lint-free tissue. If the other dropper bottles are also wet
or have droplets on their tips, these should be dried as well with great care taken not
to cross-contaminate the tips by wiping them with the same section of tissue that has
already been used to wipe another dropper bottle tip. Dropper bottle tips can become
wet if the bottles are stored on their sides or turned upside down. Ideal storage is vertical
configuration of all dropper bottles.
ƒƒ Suspend the dropper bottle at a 45° angle so that the tip is just above the centre of the
foam pad wells, without touching the foam pad itself. The drop should be targeted to freefall into the centre of the well. Do not rest the dropper bottle on the side of the foam pad
well when dispensing the drops as this can lead to the drops being drawn back onto the
foam pad instead of into the well.
ƒƒ At no time should any of the bottle dropper tips come into contact with the skin at the
base of the foam pad wells or anywhere on the palm.
ƒƒ If any liquid ends up on the foam pad, dry it off gently with a clean absorbent tissue.
ƒƒ Visually confirm throughout all stages of the test that no reagent leakage has occurred
around the base of the foam pad where it meets the skin. If reagent leakage occurs
discontinue the test immediately and repeat it on the alternate palm, or wait 24 hours and
test the same palm once again.
28
User Manual September 13 Final Version.indd 28
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
ƒƒ Reagent leakage seldom occurs and can be the result of a patient’s excess perspiration,
the insufficient washing of the palm prior to testing, or the application of the foam pad
to a palm that was not yet completely dry from the use of the alcohol swab. If reagent
leakage persists on the patient’s alternate palm, a correct reading may not be possible.
ƒƒ It is important for the consistency of test results that full drops of reagent be cleanly
deposited into the foam pad test wells and that the presence of any air bubbles in the
centre well is minimized. Any air bubbles will need to be dissipated.
ƒƒ If you see bubbles present on top or at the bottom of the liquid in the centre well, you can
try gently tapping the side of the foam pad with your finger to dissipate the bubbles. There
must be no air bubbles to interfere with the binding of the Detector or the skin cholesterol
reading, which is taken by the Spectrophotometer when a beam of light shines straight
down out of the Optical Aperture and into the centre well and through the reagent fluid.
ƒƒ Bubbles appearing in any of the wells except the centre well, are inconsequential and
should not be of any concern.
ƒƒ If while squeezing out the drops, the dropper bottle tip appears bubbly or foamy, this may
mean you have reached the end of the dropper bottle contents and may not be able to
deploy sufficient reagent to conduct the test, or to continue it.
ƒƒ If this occurs prior to the reagent being dropped and no reagent has come into contact
with the patient’s hand, source a new bottle (at room temperature) to conduct the test. If
the same situation occurs and reagent has already partially come into contact with the
palm, and a replacement bottle is not immediately available at room temperature, the test
must be aborted as the extended incubation period due to delays will result in an INVALID
TEST result.
6.5 Blotting
ƒƒ The PreVu Test Operator must employ a consistent blotting technique as demonstrated in
the Training Video CD and the Training Video presentations at www.prevu.com. A deviation
in blotting technique from the prescribed norm can result in inaccurate skin cholesterol
readings being taken and an incorrect result.
ƒƒ The downward pressure of the blotting stick when first inserted into the centre foam pad
well for three seconds, should be firm enough to visually depress the skin to ensure all
the unbound reagent is removed, but not so hard as to cause the foam pad to detach
from the palm at the base of the foam well. This will cause liquid to become trapped
underneath the foam pad leading to an incorrect result.
ƒƒ When reversing the blotting stick and placing the other end into the centre well once
again, ensure that when twisting the blotting stick a quarter turn clockwise and then
counter-clockwise back, that care is taken not to exceed twisting the blotting stick any
further than the quarter turn prescribed.
ƒƒ Given the diameter of the blotting stick, a quarter turn is a very small twist and you
should take care in focusing your attention on the top of the blotting stick to try and best
approximate the quarter turn by visually tracking the rotation forward and backward.
Again, the twisting is intended to further assist the blotting stick’s built-in fluid draw
capability and must not be overdone, nor the blotting stick depressed too firmly while
twisting, as this too can result in removing Detector that has already bound to the skin.
www.prevu.com
User Manual September 13 Final Version.indd 29
1-888-928-7205 (Canada)
29
12-09-13 1:34 PM
ƒƒ When blotting the diamond shaped well, move the blotting stick from corner to corner
inside the well while maintaining contact with the skin to ensure all the excess Positive
Control Reagent is removed. The square shaped well which is the Negative Control, is
never dried by the blotting stick.
ƒƒ Practice blotting on your own hand to get the feel for firm, but not undue pressure.
Again, the goal is to dry up any excess fluid in the well and the blotting sticks have
been designed to instantly draw up the fluid. The depression and twisting of the blotting
stick therefore is intended to assist with the fluid drawing action, but must not result
in removing reagent that has already bound to the skin surface by being too firmly
depressed. This can lead to inaccurate test results being taken and an incorrect result.
6.6 Reading the Skin Cholesterol Level
ƒƒ Prior to reading the skin cholesterol level, visually confirm that no reagent leakage has
occurred around the base of the foam pad where it meets the skin. If reagent leakage
occurs discontinue the test immediately and repeat it on the alternate palm, or wait 24
hours and test the same palm once again.
ƒƒ The skin cholesterol reading must be taken with the PreVu Handheld Spectrophotometer
approaching the hypothenar eminence where the foam pad sits, from off of the edge of
the bottom of the palm. The reading must never be taken with the Spectrophotometer
crossing over the full palm of the hand from the thumb side down as this will result in a
poor fit of the Alignment Shoe along with an inability to hold the Spectrophotometer level,
resulting in an Invalid Test result. The images below demonstrate both the correct and
incorrect way to position the Spectrophotometer while taking the skin
cholesterol measurement.
Correct Reading
Incorrect Reading
ƒƒ When preparing for and taking the reading, you should place your non-dominant hand
beneath the patient’s hand for support, and maintain the palm as straight out and level as
possible as the reader is depressed, or alternatively, ensure the palm is properly balanced
resting directly on the table.
ƒƒ You should be ready to position the Spectrophotometer over the palm prior to the reading
taking place (you will have ten seconds to do this after the beep prompts you), taking
care to fit the Alignment Shoe precisely over top of and down onto the foam pad, without
causing the Alignment Shoe to close and without disturbing the fluid in the foam
pad wells.
30
User Manual September 13 Final Version.indd 30
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
ƒƒ You will notice that the Alignment Shoe features an inset space at its leading, clear plastic
edge designed to receive the foam pad, and that it is the same size and depth of the foam
pad with a hole to be targeted directly over the centre well. This recess is where the foam
pad retreats into when the Alignment Shoe is placed snugly over top of it and where it
remains during the skin cholesterol reading.
ƒƒ When it is time to take the reading, and with the Alignment Shoe retaining its precise fit
over the foam pad, you will gently press the Spectrophotometer down towards the palm
closing the shoe in the process. This needs to be done while ensuring that the Alignment
Shoe remains as perfectly level as possible to the palm and the foam pad throughout, and
with the palm remaining level as well.
ƒƒ You will notice that the Spectrophotometer is ergonomically designed with a slight bend
at the section where the Alignment Shoe is attached. This allows the hand to more
comfortably hold the device. When the Alignment Shoe is fitted to the foam pad and the
skin cholesterol reading is in process, it is this top portion of the Spectrophotometer that
must remain level with the foam pad and the palm – the bottom portion of the device will
naturally be pointed downward slightly and off the palm as a result.
ƒƒ You must take care not to press the Spectrophotometer down too hard onto the palm
while the skin cholesterol reading is taking place. While some slight compression of
the foam pad may occur, care should be taken not to compress the foam pad such that
fluid could escape the wells and contaminate the reading. A gentle pressing down of the
Spectrophotometer so that the Alignment Shoe closes completely onto the palm of the
patient’s hand during the skin cholesterol reading is all that is required.
ƒƒ Some wetness of the Alignment Shoe and of the top of the foam pad around the wells
post-reading is not abnormal and will occasionally occur.
ƒƒ Sometimes while cleaning a patient’s hand or the Alignment Shoe post-test, your hands
may come into contact with reagent from the just conducted test. If this occurs, you
should wash your hands thoroughly or wipe them with a Wet Nap® moist towelette to
avoid contaminating the next test when handling the new patient’s palm, or the foam pad
and blotting stick.
www.prevu.com
User Manual September 13 Final Version.indd 31
1-888-928-7205 (Canada)
31
12-09-13 1:34 PM
SECTION 7
PreVu Test Operator Training
And Qualification
Before testing actual patients, it is the responsibility of all persons conducting the PreVu NonInvasive Skin Cholesterol Point of Care (POC) Test to ensure they have successfully completed the
prescribed training and are qualified to test. Miraculins Inc. has established a protocol to formally
train PreVu Test Operators and to address qualification through www.prevu.com.
PreVu Test Operators must demonstrate that they are able to obtain correct Skin Cholesterol
results, as determined by comparison to established reference values.
7.1 Training and Qualification Procedure
7.1.1 Training Review
Read and ensure that you understand all sections of this PreVu Handheld Spectrophotometer
User Manual and view the Training Video CD included with the PreVu Handheld
Spectrophotometer or the Training Video presentation at www.prevu.com under the PreVu Test
Operator Training and Qualification section.
7.1.2 Practicing the Test
Practice conducting the test on a sufficient number of volunteers utilizing the reagents and test
materials provided in the PreVu Test Operator Training Kit, until you are comfortable with the
testing technique. Test volunteers must be advised that the skin cholesterol readings generated
are strictly for demonstration and practice purposes only and may not represent accurate results.
7.1.3 Qualification Testing
For final qualification, place Spectrophotometer into Qualification Mode to access numerical
patient test scores (see Section 3.8.4: Turning on the Qualification Mode), test three final
volunteers (not actual patients) on both palms, and record the numerical results (for convenience)
on the Test Sheet that can be printed out at www.prevu.com. Test volunteers must be advised
that the skin cholesterol readings generated are strictly for demonstration and practice purposes
only and may not represent accurate results.
Please note that numerical test scores cannot be accessed afterwards under REVIEW, so PreVu
Test Operators are advised to record test volunteer scores on the Test Sheet provided at the time
of testing for each of the three final test volunteers. In REVIEW test results are only available in
category format: Normal, Elevated or Borderline.
See Section 5.1: Accessing Stored Results for further information.
7.1.4 Login
Obtain a login name and password at www.prevu.com under the PreVu Test Operator Training
and Qualification section.
32
User Manual September 13 Final Version.indd 32
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
7.1.5 Submit Results
Submit the numerical results from the three qualification test subjects in the fields provided.
7.1.6 Qualification Criteria
Once the results are submitted at www.prevu.com, the PreVu Test Operator will be considered to
have successfully completed training, provided the following criteria has been met and the PreVu
Test Operator has received confirmation of same by email:
a. The coefficient of variation (CV) between the tests conducted on two palms is less
than 20%.
b. The mean skin cholesterol value for all three volunteers (six results) is less than 120.
NOTE: If the CV is greater than 20% or the mean skin cholesterol value for all three
volunteers is greater than 120, the PreVu Test Operator will be asked to review the Training
Video CD or the Training Video presentation at www.prevu.com. The PreVu Test Operator
will also be asked to go back and review the PreVu Handheld Spectrophotometer User
Manual, and practice on additional volunteers before repeating qualification testing on
three final test volunteers and submitting the data results once again.
7.1.7 Qualification Confirmation
When results satisfy the criteria outlined above, an email will be sent to the PreVu Test Operator
confirming that he/she has successfully completed training and is now qualified as being ready
to conduct the PreVu Non-Invasive Skin Cholesterol Test Point of Care (POC) Test on patients.
7.1.8 PreVu Test Operator Training and Qualification Summary
The PreVu Test Operator should confirm that he/she:
ƒƒ has thoroughly reviewed and understands the training materials for the PreVu NonInvasive Skin Cholesterol Point of Care (POC) Test;
ƒƒ has tested a sufficient number of volunteers to acquire familiarity of the testing process
and to perfect the testing technique;
ƒƒ has tested and recorded three final volunteers for qualification and submitted their results
to www.prevu.com; and
ƒƒ WILL NOT start testing actual patients until they receive email confirmation from
www.prevu.com that they have successfully completed training and have been qualified
to conduct the PreVu Non-Invasive Skin Cholesterol Test Point of Care (POC) Test on
patients.
www.prevu.com
User Manual September 13 Final Version.indd 33
1-888-928-7205 (Canada)
33
12-09-13 1:34 PM
SECTION 8
Solving Problems: Warnings and
Error Messages
The following is a summary of Warnings and Error Messages that may be displayed on the PreVu
Handheld Spectrophotometer screen. In the event of a problem, please refer to the information
under the Message, What it Means, and User Action columns. If further assistance is required,
please contact PreVu Customer Service & Support at 1-888-928-7205 (Canada) or
visit www.prevu.com.
No.
Message
What it Means
User Action
1
The Spectrophotometer
temperature is less than
10ºC (50ºF). Wait until
the device reaches a
temperature within the
operating range of 10ºC
to 40ºC (50ºF to 104ºF).
Wait until the
Spectrophotometer warms
up. The Reagent Temp
screen will appear once the
operating temperature is
reached.
2
The Spectrophotometer
has detected that the
temperature is out of
operating range of 10ºC to
40ºC (50ºF to 104ºF) during
the startup test or during
operation.
Turn Spectrophotometer
off and check the room
temperature. Turn the
unit back on after it
has adjusted within the
operating range. If the error
message appears again,
contact PreVu Customer
Service & Support. If it is
over 30ºC (86ºF) and the
batteries must be charged,
then charge with the
instrument turned off.
3
Hardware Error
Turn Spectrophotometer off
and then back on. If error
message appears again,
contact PreVu Customer
Service & Support.
4
Hardware Error
Turn Spectrophotometer off
and then back on. If error
message appears again,
contact PreVu Customer
Service & Support.
34
User Manual September 13 Final Version.indd 34
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
PreVu
ion
d,
No.
Message
What it Means
5
Hardware Error
Turn Spectrophotometer off
and then back on. If error
message appears again,
contact PreVu Customer
Service & Support.
6
Hardware Error
Turn Spectrophotometer off
and then back on. If error
message appears again,
contact PreVu Customer
Service & Support.
7
Hardware Error
Turn Spectrophotometer off
and then back on. If error
message appears again,
contact PreVu Customer
Service & Support.
8
Battery is too low.
Plug in the
Spectrophotometer for at
least 15 minutes or replace
the batteries with 2 fully
charged AA rechargeable
batteries (refer to section
3.3.1 and 3.3.2 of this
manual for further
instruction)
9
The battery level is too low
to perform a test.
Plug in the
Spectrophotometer for at
least 15 minutes or replace
the batteries with 2 fully
charged AA rechargeable
batteries (refer to section
3.3.1 and 3.3.2 of this
manual for further
instruction)
10
Calibration error during the
start of the new test.
Press the OK button to
return to the CALIBRATION
screen. Repeat Calibration
procedure. If the Calibration
Error message appears
again, contact PreVu
Customer Service
and Support.
rms
e the
is
er
m
e
error
n,
er
is
he
ged,
er off
ror
n,
er
User Action
er off
ror
n,
er
www.prevu.com
User Manual September 13 Final Version.indd 35
1-888-928-7205 (Canada)
35
12-09-13 1:34 PM
No.
Message
What it Means
User Action
11
Skin Cholesterol value
out of the preset Skin
Cholesterol range.
Refer to the Package Insert
of the PreVu Non-Invasive
Skin Cholesterol Point of
Care (POC) Test Reagent
Kit, to ensure the test is
performed properly, and
then repeat test.
12
Timeout (late) during the
measurement.
Repeat the test (if
conducted within 24
hours test must be
done on patient’s other
palm), ensuring that the
measurement is taken
within 10 seconds of when
the 2 minute (Indicator)
timer expires.
13
Hardware error.
Turn Spectrophotometer off
and then back on. Use the
MENU>REVIEW to see
results. If error message
appears again, contact
PreVu Customer Service
and Support.
14
Shoe opened during
measurement.
Repeat the test. Use care
to keep the shoe closed
during the measurement.
36
User Manual September 13 Final Version.indd 36
1-888-928-7205 (Canada)
www.prevu.com
12-09-13 1:34 PM
nsert
ive
of
nt
s
d
e
when
)
er off
the
SECTION 9
Spectrophotometer Support and Care
9.1 Warranty and Repair Information
For information about Instrument Warranty and Repair, please contact PreVu Customer Service
and Support at 1-888-928-7205 (Canada) or visit www.prevu.com.
9.2 Spectrophotometer Care
ƒƒ Keep the instrument clean
ƒƒ Do not immerse in water or other liquid
ƒƒ Do not put any liquid into the Optical Aperture on the bottom of the instrument
9.3 Cleaning Instructions
ƒƒ Use a cloth lightly dampened with water and mild detergent to wipe down the outside of
the instrument
ƒƒ Remove all liquid from the alignment shoe after each test
e
9.4 Alignment Shoe Replacement
e
The Alignment Shoe may be subject to wear from regular movement. If necessary, to inquire
about replacing your Alignment Shoe, please contact PreVu Customer Service and Support.
re
d
nt.
Alignment Shoe
www.prevu.com
User Manual September 13 Final Version.indd 37
1-888-928-7205 (Canada)
37
12-09-13 1:34 PM
PreVu Customer Service and Support
1-888-928-7205
www.prevu.com
PreVu Handheld Spectrophotometer
(PVU-SPECT-101-UNI)
P101-MA-025-01
This User Manual is strictly intended for use by PreVu Test Operators in Canada.
Manufactured by: Miraculins Inc., 6-1250 Waverley Street, Winnipeg, Manitoba, Canada, R3T 6C6
The PreVu® Non-Invasive Skin Cholesterol Test™ technology is protected by patents and patents pending in
North America and internationally.
PreVu® is a Registered Trademark of Miraculins Inc. All Rights Reserved. 2012.
™ Designates Property of Miraculins Inc. All Rights Reserved. 2012.
Wet-Nap® is a Registered Trademark of Nice Pak, Inc.
Copyright © 2012 by Miraculins Inc. All Rights Reserved. No part of this User Manual, including interior
design, cover design, and images, may be reproduced or transmitted in any form, by any means (electronic,
photocopying, recording, or otherwise) without the prior written permission of Miraculins Inc.
Printed in Canada.
User Manual September 13 Final Version.indd 38
12-09-13 1:34 PM